MRI is often used in tumor localization for radiotherapy treatment planning, with gadolinium (Gd)-containing materials often introduced as a contrast agent. Motexafin gadolinium is a novel radiosensitizer currently being studied in clinical trials. The nanoparticle technologies can target tumors with high concentration of high-Z materials. This Monte Carlo study is the first detailed quantitative investigation of high-Z material Gd-induced dose enhancement in megavoltage external beam photon therapy. BEAMnrc, a radiotherapy Monte Carlo simulation package, was used to calculate dose enhancement as a function of Gd concentration. Published phase space files for the TrueBeam flattening filter free (FFF) and conventional flattened 6MV photon beams were used. High dose rate (HDR) brachytherapy with Ir-192 source was also investigated as a reference. The energy spectra difference caused a dose enhancement difference between the two beams. Since the Ir-192 photons have lower energy yet, the photoelectric effect in the presence of Gd leads to even higher dose enhancement in HDR. At depth of 1.8 cm, the percent mean dose enhancement for the FFF beam was 0.38±0.12, 1.39±0.21, 2.51±0.34, 3.59±0.26, and 4.59±0.34 for Gd concentrations of 1, 5, 10, 15, and 20 mg/mL, respectively. The corresponding values for the flattened beam were 0.09±0.14, 0.50±0.28, 1.19±0.29, 1.68±0.39, and 2.34±0.24. For Ir-192 with direct contact, the enhanced were 0.50±0.14, 2.79±0.17, 5.49±0.12, 8.19±0.14, and 10.80±0.13. Gd-containing materials used in MRI as contrast agents can also potentially serve as radiosensitizers in radiotherapy. This study demonstrates that Gd can be used to enhance radiation dose in target volumes not only in HDR brachytherapy, but also in 6 MV FFF external beam radiotherapy, but higher than the currently used clinical concentration (>5 mg/mL) would be needed.  相似文献   

2.

Background

Safe High Dose Rate Brachytherapy (HDR-BT) requires quality assurance/quality control (QA/QC) according to IPEM and ESTRO recommendations. Recent advances in real-time dosimetry and related developments of QA, QC and in vivo dosimetry (IVD) systems have offered new possibilities for effective independent treatment verification, and thus for improving the patient safety.

Contributions

This paper briefly reports the state of the art of different QA and QC approaches and systems. It also presents our related studies, carried out within the framework of DoRGaN research project, which have led to two proposed instruments: a QA pretreatment instrumented phantom and a QC gynecological applicator. Both systems implement a real-time dwell-source distance-measuring approach using multiple Gallium Nitride (GaN) dosimetry probes. First prototypes have been designed and tested. Their characteristics (including those of the employed GaN probes) are evaluated in comparison with the state of the art.

Conclusion

For developing systems for HDR-BT, the GaN probe has been shown to meet requirements for high sensitivity, high measurement rate, small response anisotropy and large dose measurement range. The proposed QA instrumented phantom has comparable or better characteristics when compared with diode-based and Al2O3:C-based systems. The other QC gynecological applicator has higher time and spatial resolutions than the state-of-the-art IVD systems, which proves that it is suitable for real-time monitoring of HDR-BT physical parameters.  相似文献   

3.
60Co sources are being used as an alternative to 192Ir sources in high dose rate brachytherapy treatments. In a recent document from AAPM and ESTRO, a consensus dataset for the 60Co BEBIG (model Co0.A86) high dose rate source was prepared by using results taken from different publications due to discrepancies observed among them. The aim of the present work is to provide a new calculation of the dosimetric characteristics of that 60Co source according to the recommendations of the AAPM and ESTRO report. Radial dose function, anisotropy function, air-kerma strength, dose rate constant and absorbed dose rate in water have been calculated and compared to the results of previous works. Simulations using the two different geometries considered by other authors have been carried out and the effect of the cable density and length has been studied.  相似文献   

4.

Aim

Stepping source in brachytherapy systems is used to treat a target lesion longer than the effective treatment length of the source. Cancerous lesions in the cervix, esophagus and rectum are examples of such a target lesion.

Background

In this study, the stepping source of a GZP6 afterloading intracavitary brachytherapy unit was simulated using Monte Carlo (MC) simulation and the results were used for the validation of the GZP6 treatment planning system (TPS).

Materials and methods

The stepping source was simulated using MCNPX Monte Carlo code. Dose distributions in the longitudinal plane were obtained by using a matrix shift method for esophageal tumor lengths of 8 and 10 cm. A mesh tally has been employed for the absorbed dose calculation in a cylindrical water phantom. A total of 5 × 108 photon histories were scored and the MC statistical error obtained was at the range of 0.008–3.5%, an average of 0.2%.

Results

The acquired MC and TPS isodose curves were compared and it was shown that the dose distributions in the longitudinal plane were relatively coincidental. In the transverse direction, a maximum dose difference of 7% and 5% was observed for tumor lengths of 8 and 10 cm, respectively.

Conclusion

Considering that the certified source activity is given with ±10% uncertainty, the obtained difference is reasonable. It can be concluded that the accuracy of the dose distributions produced by GZP6 TPS for the stepping source is acceptable for its clinical applications.  相似文献   

5.

Background

As a routine method for stepping source simulation, a Monte Carlo program is run according to the number of steps and then the summation of dose from each run is taken to obtain total dose distribution. This method is time consuming.

Aim

As an alternative method, a matrix shift based technique was applied to simulate a stepping source for brachytherapy.

Materials and methods

The stepping source of GZP6 brachytherapy unit was simulated. In a matrix shift method, it is assumed that a radiation source is stationary and instead the data matrix is shifted based on the number of steps. In this study, by running MCNPX program for one point and calculation of the dose matrix using the matrix shift method, the isodose curves for the esophageal cancer tumor lengths of 4 and 6 cm were obtained and compared with the isodose curves obtained by running MCNPX programs in each step position separately (15 and 23 steps for esophageal cancer tumor lengths of 4 and 6 cm, respectively).

Results

The difference between the two dose matrixes for the stepping and matrix shift methods based on the average dose differences are 3.85 × 10−4 Gy and 5.19 × 10−4 Gy for treatment length of 4 cm and 6 cm, respectively. Dose differences are insignificant and these two methods are equally valid.

Conclusions

The matrix shift method presented in this study can be used for calculation of dose distribution for a brachytherapy stepping source as a quicker tool compared to other routine Monte Carlo based methods.  相似文献   

6.
Detailed-history Monte Carlo computer codes were used to simulate the formation, diffusion, and chemical reaction of free-radical species within deoxygenated aqueous solutions of glycylglycine irradiated by 250-kVp X rays and by 60Co gamma rays. In one reaction, hydrated electrons react with the glycylglycine solute to produce unbound, or free, ammonia. This reaction is complete by 10(-6) s within individual electron tracks for glycylglycine concentrations greater than or equal to 0.025 M. For solute concentrations from 0.025 to 1.2 M, calculated G values of free ammonia are in excellent agreement with measured values. In addition, the computer model predicts a statistically significant difference between the G value of free ammonia produced under X irradiation and that produced under 60Co gamma irradiation.  相似文献   

7.
The Hardy-Weinberg law is among the most important principles in the study of biological systems. Given its importance, many tests have been devised to determine whether a finite population follows Hardy-Weinberg proportions. Because asymptotic tests can fail, Guo and Thompson developed an exact test; unfortunately, the Monte Carlo method they proposed to evaluate their test has a running time that grows linearly in the size of the population N. Here, we propose a new algorithm whose expected running time is linear in the size of the table produced, and completely independent of N. In practice, this new algorithm can be considerably faster than the original method.  相似文献   

8.
A Monte Carlo computer simulation program is designed in orderto describe the spatial and time evolution of a population ofliving individuals under preassigned environmental conditionsof energy. The simulation is inspired by previous techniquesdeveloped in physics — in particular, in molecular dynamicsand simulations of liquids — and it already provides somenew insights regarding macroscopic deterministic models in ecologyand concerning eventual control of artificial biomass productionplants. Received on July 15, 1986; accepted on October 9, 1986  相似文献   

9.
Bioluminescence imaging plays an important role in the areas of cancer biology, cell biology, gene therapy, and so on. The 2D planar bioluminescent imaging has been transformed into a 3D framework by bioluminescence tomography (BLT) that enables bioluminescent source reconstruction in a mouse using a modality fusion approach. To solve this BLT problem, a geometrical model of the mouse is usually built from a CT/micro-CT/micro-MRI scan, which facilitates the assignment of optical parameters to various anatomical regions in the model. This optical model is then used to facilitate BLT. The forward model is based on Monte Carlo simulation to calculate the diffuse light flux on the surface of the mouse. The forward model data are used to define the imaging system and perform the BLT reconstruction. In this paper, we report the reconstruction of sources inside a heterogeneous highly scattering physical phantom to demonstrate the feasibility of this Monte Carlo based BLT method.  相似文献   

10.
Cobalt 60 source is generally available on high dose rate (HDR) afterloading equipment especially for treatment of gynecological lesions. The GZP6 remote afterloader (Nuclear Power Institute of China) utilizes 60Co sources for treatment of intracavitary and intraluminal malignancies. In this study, the AAPM TG-43 dosimetric parameters of three sources in GZP6 system have been studied using MCNP4C Monte Carlo (MC) code; and the results are compared with other available 60Co HDR sources. The presented parameters consist of air kerma strength, dose rate constant, radial dose function and anisotropy function. They show less than 1% uncertainty. The TG-43 based dosimetry data can be used not only to validate the dedicated treatment planning software (TPS), but also to introduce new complementary software to enhance the system performance in gynecological treatments.  相似文献   

11.
12.
13.
14.
Monte Carlo feature selection for supervised classification   总被引:4,自引:0,他引:4  
MOTIVATION: Pre-selection of informative features for supervised classification is a crucial, albeit delicate, task. It is desirable that feature selection provides the features that contribute most to the classification task per se and which should therefore be used by any classifier later used to produce classification rules. In this article, a conceptually simple but computer-intensive approach to this task is proposed. The reliability of the approach rests on multiple construction of a tree classifier for many training sets randomly chosen from the original sample set, where samples in each training set consist of only a fraction of all of the observed features. RESULTS: The resulting ranking of features may then be used to advantage for classification via a classifier of any type. The approach was validated using Golub et al. leukemia data and the Alizadeh et al. lymphoma data. Not surprisingly, we obtained a significantly different list of genes. Biological interpretation of the genes selected by our method showed that several of them are involved in precursors to different types of leukemia and lymphoma rather than being genes that are common to several forms of cancers, which is the case for the other methods. AVAILABILITY: Prototype available upon request.  相似文献   

15.
16.
The purpose of this note is to illustrate the feasibility of simulating kinetic systems, such as commonly encountered in photosynthesis research, using the Monte Carlo (MC) method. In this approach, chemical events are considered at the molecular level where they occur randomly and the macroscopic kinetic evolution results from averaging a large number of such events. Their repeated simulation is easily accomplished using digital computing. It is shown that the MC approach is well suited to the capabilities and resources of modern microcomputers. A software package is briefly described and discussed, allowing a simple programming of any kinetic model system and its resolution. The execution is reasonably fast and accurate; it is not subject to such instabilities as found with the conventional analytical approach.Abbreviations MC Monte Carlo - RN random number - PSU photosynthetic unit Dedicated to Prof. L.N.M. Duysens on the occasion of his retirement.  相似文献   

17.
This study explores how the metal materials of the applicator influence the dose distribution when performing brachytherapy for cervical cancer. A pinpoint ionization chamber, Monte Carlo code MCNPX, and treatment planning system are used to evaluate the dose distribution for a single Ir-192 source positioned in the tandem and ovoid. For dose distribution in water with the presence of the tandem, differences among measurement, MCNPX calculation and treatment planning system results are <5%. For dose distribution in water with the presence of the ovoid, the MCNPX result agrees with the measurement. But the doses calculated from treatment planning system are overestimated by up to a factor of 4. This is due to the shielding effect of the metal materials in the applicator not being considered in the treatment planning system. This result suggests that the treatment planning system should take into account corrections for the metal materials of the applicator in order to improve the accuracy of the radiation dose delivered.  相似文献   

18.
19.
BackgroundThe presence of heterogeneity within the radiation field increases the challenges of small field dosimetry. In this study, the performance of MAGIC polymer gel was evaluated in the dosimetry of small fields beyond bone heterogeneity.Materials and methodsCircular field sizes of 5, 10, 20 and 30 mm were used and Polytetrafluoroethylene with density of 2.2 g/cm3 was used as the bone equivalent material. The PDD curves, beam profiles, and penumbra widths were measured using MAGIC polymer gel, EBT2 film, and Monte Carlo simulation.ResultsThe maximum differences between MAGIC and EBT2 are 6.1, 4.7, 2.4, and 2.2 for PDD curves at 5, 10, 20, and 30 mm circular fields, respectively. The dose differences and distance to agreement between MAGIC and MC were within 1.89%/0.46 mm, 1.66%/0.43 mm, 1.28%/0.77 mm, and 1.31%/0.81 mm for beam profile values behind bone heterogeneity at 5, 10, 20, and 30 mm field sizes, respectively.ConclusionThe results presented that the MAGIC polymer gel dosimeter is a proper instrument for dosimetry beyond high density heterogeneity.  相似文献   