首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Amperometric biosensor for hydrogen peroxide based on ferrocene-bovine serum albumin and multiwall carbon nanotube modified ormosil composite
Authors:Tripathi Vijay Shyam  Kandimalla Vivek Babu  Ju Huangxian
Institution:Key Laboratory of Analytical Chemistry for Life Science (Education Ministry of China), Department of Chemistry, Nanjing University.
Abstract:A novel amperometric biosensor for hydrogen peroxide (H(2)O(2)) was developed by entrapping horseradish peroxidase (HRP) in a new ormosil composite doped with ferrocene monocarboxylic acid-bovine serum albumin conjugate and multiwall carbon nanotubes (MWNTs). The ormosil was prepared using 3-(aminopropyl)triethoxysilane and 2-(3,4 epoxycyclohexyl)-ethyltrimethoxy silane as monomers. The encapsulated conjugate showed excellent electrochemistry and acted as an electron transfer mediator. The presence of MWNTs improved the conductivity of the composite film. This matrix showed a biocompatible microenvironment for retaining the native activity of the entrapped HRP and a very low mass transport barrier to the substrate, which provided a fast amperometric response to H(2)O(2). The proposed H(2)O(2) biosensor exhibited a linear range of 0.02-4.0 mM with a detection limit of 5.0 microM (S/N = 3) and a K(M)(app) value of 2.0 mM. It could be used for flow injection analysis of hydrogen peroxide with a liner range from 0.02 to 4.5 mM, sensitivity of 0.042 microA/mM and analytical time of 20 s per sample. This biosensor possessed good analytical performance and storage stability.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏