EGSnrc application for IMRT planning
Institution:1. Department of Physics, Faculty of Mathematics and Natural Sciences, Bogor Agricultural University (IPB University), Jalan Meranti Kampus IPB Dramaga, Bogor 16680, Indonesia;2. Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung, Jalan Ganesa 10, Coblong, Bandung, West Java, 40132, Indonesia;3. Indonesian National Institute of Aeronautics and Space, Jl. Goa Jepang, Sumberker, Samofa, Kabupaten Biak Numfor, Papua 98118, Indonesia;4. Radiolology Department, National University Cancer Institute, Singapore, 119074, Singapore;1. Department of Physics, Faculty of Mathematics and Natural Sciences, Bogor Agricultural University (IPB University), Jalan Meranti Kampus IPB Dramaga, Bogor 16680, Indonesia;2. Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung, Jalan Ganesa 10, Coblong, Bandung, West Java, 40132, Indonesia;3. Indonesian National Institute of Aeronautics and Space, Jl. Goa Jepang, Sumberker, Samofa, Kabupaten Biak Numfor, Papua 98118, Indonesia;4. Radiolology Department, National University Cancer Institute, Singapore, 119074, Singapore
Abstract:The aim of this study was to describe a detailed instruction of intensity modulated radiotherapy (IMRT) planning simulation using BEAMnrc-DOSXYZnrc code system (EGSnrc package) and present a new graphical user interface based on MATLAB code (The MathWorks) to combine more than one. 3ddose file which were obtained from the IMRT plan.This study was performed in four phases: the commissioning of Varian Clinac iX6 MV, the simulation of IMRT planning in EGSnrc, the creation of in-house VDOSE GUI, and the analysis of the isodose contour and dose volume histogram (DVH) curve from several beam angles. The plan paramaters in sequence and control point files were extracted from the planning data in Tan Tock Seng Hospital Singapore (multileaf collimator (MLC) leaf positions – bank A and bank B, gantry angles, coordinate of isocenters, and MU indexes).VDOSE GUI which was created in this study can display the distribution dose curve in each slice and beam angle. Dose distributions from various MLC settings and beam angles yield different dose distributions even though they used the same number of simulated particles. This was due to the differences in the MLC leaf openings in every field. The value of the relative dose error between the two dose ditributions for “body” was 51.23 %. The Monte Carlo (MC) data was normalized with the maximum dose but the analytical anisotropic algorithm (AAA) data was normalized by the dose in the isocenter.In this study, we have presented a Monte Carlo simulation framework for IMRT dose calculation using DOSXYZnrc source 21. Further studies are needed in conducting IMRT simulations using EGSnrc to minimize the different dose error and dose volume histogram deviation.
Keywords:Monte carlo  EGSnrc  IMRT planning  VDOSE
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号