首页 | 本学科首页   官方微博 | 高级检索  
   检索      


HPLC fluorescence method for the determination of nizatidine in human plasma and its application to pharmacokinetic study
Authors:Mahmut B Çakar  Sevgi T Ulu
Institution:Istanbul University, Faculty of Pharmacy, Department of Analytical Chemistry, Istanbul, Turkey
Abstract:A sensitive high‐performance liquid chromatographic (HPLC) method was developed for the determination of nizatidine in human plasma. Nizatidine was derivatized by 4‐fluoro‐7‐nitrobenzofurazan (NBD‐F). Chromatographic separation was performed on a Inertsil C18 column (150 mm × 4.6 mm, 5 µm) using isocratic elution by a mobile phase consisting of methanol/water (55:45) at a flow rate of 1.2 mL/min. Amlodipine was used as the internal standard (IS). Fluorescence detector was used operated at 461 nm (excitation) and 517 nm (emission), respectively. The calibration curve was linear over the range of 50–2000 ng/mL. This method was successfully applied to a pharmacokinetic study after oral administration of a dose (150 mg) of nizatidine. Copyright © 2013 John Wiley & Sons, Ltd.
Keywords:nizatidine  derivatization  pharmacokinetics  fluorescence detector