首页 | 本学科首页   官方微博 | 高级检索  
   检索      


A PVC-based electrode sensitive to DDA+ as a device for monitoring the membrane potential in biological systems
Authors:T Shinbo  N Kamo  K Kurihara  Y Kobatake
Institution:Faculty of Pharmaceutical Sciences, Hokkaido University, Sapporo, Japan
Abstract:A polyvinyl chloride (PVC)-based membrane electrode sensitive to dibenzyldimethyl ammonium cation (DDA+) was constructed, and operational parameters such as the selectivity coefficients, the detection limit, and the response time were obtained. In comparison with the selectivity coefficients obtained with the previous liquid-membrane electrode, significant improvement was not obtained, but the response time became pronouncedly shorter. Furthermore, the electrode lifetime was remarkably prolonged. With the electrode developed, the change in the membrane potential of liposomes containing dibutyl ferrocene which separated oxidizing and reducing agent solutions was measured. The DDA+ uptake, U, and the membrane potential estimated from U changed in accordance with the redox potential in the medium when the concentration of internal ferricyanide was kept constant. The membrane potential collapsed when the uncoupler of oxidative phosphorylation was added. The ANS fluorescence measurement indicated that negative charges appeared on energization with oxidizing-reducing agent. The change in membrane potential of mitochondria during energization was also measured. It was found that the liposome described above is a good model for the generation of membrane potentials in mitochondria.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏