首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Construction of an amperometric TG biosensor based on AuPPy nanocomposite and poly (indole-5-carboxylic acid) modified Au electrode
Authors:Jagriti Narang  Nidhi Chauhan  Poonam Rani  C S Pundir
Institution:1. Department of Biochemistry, M. D. University, Rohtak, 124 001, Haryana, India
Abstract:A method is described for construction of an amperometric triglyceride (TG) biosensor based on covalent co-immobilization of lipase, glycerol kinase and glycerol-3-phosphate oxidase onto gold polypyrrole nanocomposite decorated poly indole-5-carboxylic acid electrodeposited on the surface of a gold electrode. The enzyme electrode was characterized by transmission electron microscopy, scanning electron microscopy, electrochemical impedance studies, Fourier transform infrared spectroscopy and cyclic voltammetry. Biosensor showed optimum response within 4 s at pH 6.5 and 35 °C, when polarized at +0.1 V against Ag/AgCl. There was a linear relationship between sensor response and triolein concentration in the range 50–700 mg/dl. Biosensor was employed for determination of TG in serum. Detection limit of the biosensor was 20 mg/dl. Biosensor was evaluated with 91–95 % recovery of added triolein in sera and 4.14 and 5.85 % within and between batch coefficients of variation, respectively. There was a good correlation (r = 0.99) between sera TG values by standard method (Enzymic colorimetric) and the present method. The biosensor was unaffected by a number of serum substances at their physiological concentration. Biosensor lost 50 % of its initial activity after its 100 uses over 7 months, when stored at 4 °C.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏