CyberKnife beam output factor measurements: A multi-site and multi-detector study
Institution:1. Department of Medical Physics, Greater Poland Cancer Centre, Poznan, Poland;2. Department of Electroradiology, Poznan University of Medical Sciences, Poznan, Poland;3. Department of Technical Physics, Poznan University of Technology, Poznan, Poland
Abstract:PurposeNew promising detectors are available for measuring small field size output factors (OFs). This study focused on a multicenter evaluation of two new generation detectors for OF measurements on CyberKnife systems.MethodsPTW-60019 microDiamond and W1 plastic scintillation detector (PSD) were used to measure OFs on eight CyberKnife units of various generations for 5–60 mm fixed cones. MicroDiamond and PSD OF were compared to routinely used silicon diodes data corrected applying published Monte Carlo (MC) factors. PSD data were corrected for ?erenkov Light Ratio (CLR). The uncertainties related to CLR determination were estimated.ResultsConsidering OF values averaged over all centers, the differences between MC corrected diode and the other two detectors were within 1.5%. MicroDiamond exhibited an over-response of 1.3% at 7.5 mm and a trend inversion at 5 mm with a difference of 0.2%. This behavior was consistent among the different units. OFs measured by PSD slightly under-responded compared to MC corrected diode for the smaller cones and the differences were within 1%. The observed CLR variability was 2.5% and the related variation in OF values was 1.9%.ConclusionThis study indicates that CyberKnife microDiamond OF require corrections below 2%. The results are enhanced by the consistency observed among different units. Scintillator shows a good agreement to MC corrected diode but CLR determination remains critical requiring further investigations. The results emphasized the value of a multi-center validation over a single center approach.
Keywords:PTW-60019 microDiamond detector  W1 scintillation detector  CyberKnife output factors  Multicenter evaluation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号