首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Impact of transverse magnetic fields on dose response of a nanoDot OSLD in megavoltage photon beams
Institution:2. Department of Health Sciences, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan
Abstract:PurposeWe investigated the impact of transverse magnetic fields on the dose response of a nanoDot optically stimulated luminescence dosimetry (OSLD) in megavoltage photon beams.MethodsThe nanoDot OSLD response was calculated via Monte Carlo (MC) simulations. The responses RQ and RQ,B without and with the transverse magnetic fields of 0.35–3 T were analyzed as a function of depth at a 10 cm × 10 cm field for 4–18 MV photons in a solid water phantom. All responses were determined based on comparisons with the response under the reference conditions (depth of 10 cm and a 10 cm × 10 cm field) for 6 MV without the magnetic field. In addition, the influence of air-gaps on the nanoDot response in the magnetic field was estimated according to Burlin’s general cavity theory.ResultsThe RQ as a function of depth for 4–18 MV ranged from 1.013 to 0.993, excepting the buildup region. The RQ,B increased from 2.8% to 1.5% at 1.5 T and decreased from 3.0% to 1.1% at 3 T in comparison with RQ as the photon energy increased. The depth dependence of RQ,B was less than 1%, excepting the buildup region. The top air-gap and the bottom air- gap were responsible for the response reduction and the response increase, respectively.ConclusionsThe response RQ,B varied depending on the magnetic field intensity, and the variation of RQ,B reduced as the photon beam energy increased. The air-gaps affected the dose deposition in the magnetic fields.
Keywords:Optically stimulated luminescence dosimeters  MRI-guided radiation therapy  Monte Carlo simulation  Burlin’s general cavity theory
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号