A simple and rapid determination of biapenem in plasma by high-performance liquid chromatography |
| |
Authors: | Ikeda Kayo Ikawa Kazuro Ikeda Aki Nishikawa Yoshimi Morikawa Norifumi |
| |
Affiliation: | Department of Clinical Pharmacotherapy, Division of Clinical Pharmacotherapeutics, Graduate School of Biomedical Sciences, Hiroshima University, Kasumi 1-2-3, Minami-ku, Hiroshima 734-8553, Japan. ikeda@hiroshima-u.ac.jp |
| |
Abstract: | A simple, rapid and precise HPLC method using ultrafiltration to remove plasma protein was developed to determine biapenem concentrations in human plasma. Plasma was separated by centrifugation at 4 degrees C from blood collected in heparinized vacuum tubes, and biapenem was stabilized by immediate mixing the plasma with 1M 3-morpholinopropanesulfonic acid (MOPS) buffer (pH 7.0) (1:1). Biapenem was detected by ultraviolet absorbance at 300nm with no interfering plasma peak. The calibration curve of biapenem in human plasma was linear from 0.04 to 50microg/mL. The limit of detection was 0.01microg/mL, which was more than 40-fold lower than that of conventional plasma protein precipitation using ammonium sulfate. The assay has been clinically applied to pharmacokinetic studies in patients. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|