Miniaturized all fiber probe for optical coherence tomography and pH detection of biological tissue
Authors:Minghui Chen  Jianping Wang  Weijie Tan  Yuanyuan Feng  Gang Zheng
Institution:Shanghai Institute for Interventional Medical Devices, University of Shanghai for Science and Technology, Shanghai, China
Abstract:We present a novel all-fiber probe with 710-μm outside diameter for combined optical coherence tomography and pH detection. In cancer surgery, a significant challenge is how to completely remove the malignant tumor without cutting too much normal tissue. The difference between cancer tissue and normal tissue not only lies in morphology and structure but also in tissue pH, where malignant tissue has a lower pH. This dual-modality probe combined optical coherence tomography and pH detection of biological tissue, is expected to determine whether the tissue is cancerous quickly and accurately. The probe utilizes a typical three-segment structure (double-clad fiber - no-core fiber - graded-index fiber). We obtained a lateral resolution of ~10.6 μm, a working distance of ~506 μm and a pH measurement accuracy of 0.01 pH unit for the probe. The performance of the all-fiber probe was verified through an ex vivo experiment using the porcine brain specimen.image
Keywords:endoscopic probe  fluorescence ratio imaging  multimodal imaging  optical coherence tomography  pH measurement