Haemocompatibility of coronary catheters. |
| |
Authors: | C Mrowietz R Bach U Seyfert J W Park R P Franke F Jung |
| |
Affiliation: | Dresdner Institut für Herz- und Kreislaufforschung, Dresden. |
| |
Abstract: | BACKGROUND: Unwanted side-effects like thrombotic-thromboembolic processes accompany the intravascular use of catheters. The influence of the physico-chemical properties of the catheter materials on these processes is not completely understood. METHODS: The haemocompatibility of angiography-catheters (n = 10) (Super Torque Plus, CORDIS, Johnson & Johnson company) was examined in a perfusion-model filled with platelet-rich plasma and either compared to a non-thrombogenic (silicon-tube) or to a thrombogenic surface (glass). As test-parameters the number of single circulating thrombocytes, of circulating platelet-aggregates, of circulating activated thrombocytes (CD62) and the thrombin generation (thrombin antithrombin III-complexes) were evaluated. RESULTS: None of the test-parameters showed differences between the non-thrombogenic control-system and the angiography-catheters. Both systems, however, differed significantly and relevantly from the system filled with the glass-beads (the thrombogenic control-systems). CONCLUSION: With respect to the platelet reactivity angiography-catheters of the examined type can therefore be regarded as non-thrombogenic or haemocompatible. |
| |
Keywords: | |
|
|