首页 | 本学科首页   官方微博 | 高级检索  
     


Minimizing pulling geometry errors in atomic force microscope single molecule force spectroscopy
Authors:Rivera Monica  Lee Whasil  Ke Changhong  Marszalek Piotr E  Cole Daniel G  Clark Robert L
Affiliation:* Department of Mechanical Engineering and Materials Science, Center for Biologically Inspired Materials and Materials Systems, Pratt School of Engineering, Duke University, Durham, North Carolina 27708
Department of Mechanical Engineering, State University of New York at Binghamton, Binghamton, New York 13902
Department of Mechanical Engineering and Materials Science, University of Pittsburgh, Pittsburgh, Pennsylvania 15261
Abstract:In atomic force microscopy-based single molecule force spectroscopy (AFM-SMFS), it is assumed that the pulling angle is negligible and that the force applied to the molecule is equivalent to the force measured by the instrument. Recent studies, however, have indicated that the pulling geometry errors can drastically alter the measured force-extension relationship of molecules. Here we describe a software-based alignment method that repositions the cantilever such that it is located directly above the molecule's substrate attachment site. By aligning the applied force with the measurement axis, the molecule is no longer undergoing combined loading, and the full force can be measured by the cantilever. Simulations and experimental results verify the ability of the alignment program to minimize pulling geometry errors in AFM-SMFS studies.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!