Abstract: | Thin layer chromatography (TLC) methods for identifying and quantifying deoxynivalenol (DON), fumonisin B1 (FB1) and zearalenone in grain samples were compared to immunoassay (ELISA) and high performance liquid chromatography (HPLC) methods to determine the reliability of the less expensive TLC. There was a very good agreement between levels of DON measured by TLC and competitive-direct ELISA, and between levels of fumonisin B1 measured by TLC and HPLC, over a wide range of concentrations. Correlation coefficients (Pearson's) were 0.978, 0.914 and 0.953 for DON in maize, DON in wheat and FB1 in maize respectively. A lower correlation coefficient (r = 0.672) was obtained when zearalenone was quantified by TLC and HPLC. Possible reasons for this are discussed. A cost comparison of the various methods revealed that TLC was the least expensive for sample analysis. It is recommended that researchers choose which analytical method to use based upon individual considerations of cost and precision. This revised version was published online in June 2006 with corrections to the Cover Date. |