首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Ablation of Myocardial Tissue With Nanosecond Pulsed Electric Fields
Authors:Fei Xie  Frency Varghese  Andrei G Pakhomov  Iurii Semenov  Shu Xiao  Jonathan Philpott  Christian Zemlin
Institution:1. Department of Electrical and Computer Engineering, Old Dominion University, Norfolk, Virginia, United States of America.; 2. Center for Bioelectrics, Old Dominion University, Norfolk, Virginia, United States of America.; 3. Department of Surgery, Eastern Virginia Medical School, Norfolk, Virginia, United States of America.; Rutgers University -New Jersey Medical School, UNITED STATES,
Abstract:

Background

Ablation of cardiac tissue is an essential tool for the treatment of arrhythmias, particularly of atrial fibrillation, atrial flutter, and ventricular tachycardia. Current ablation technologies suffer from substantial recurrence rates, thermal side effects, and long procedure times. We demonstrate that ablation with nanosecond pulsed electric fields (nsPEFs) can potentially overcome these limitations.

Methods

We used optical mapping to monitor electrical activity in Langendorff-perfused New Zealand rabbit hearts (n = 12). We repeatedly inserted two shock electrodes, spaced 2–4 mm apart, into the ventricles (through the entire wall) and applied nanosecond pulsed electric fields (nsPEF) (5–20 kV/cm, 350 ns duration, at varying pulse numbers and frequencies) to create linear lesions of 12–18 mm length. Hearts were stained either with tetrazolium chloride (TTC) or propidium iodide (PI) to determine the extent of ablation. Some stained lesions were sectioned to obtain the three-dimensional geometry of the ablated volume.

Results

In all animals (12/12), we were able to create nonconducting lesions with less than 2 seconds of nsPEF application per site and minimal heating (< 0.2°C) of the tissue. The geometry of the ablated volume was smoother and more uniform throughout the wall than typical for RF ablation. The width of the lesions could be controlled up to 6 mm via the electrode spacing and the shock parameters.

Conclusions

Ablation with nsPEFs is a promising alternative to radiofrequency (RF) ablation of AF. It may dramatically reduce procedure times and produce more consistent lesion thickness than RF ablation.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏