Selective and sensitive biosensor for theophylline based on xanthine oxidase electrode |
| |
Authors: | Stredansky M Pizzariello A Miertus S Svorc J |
| |
Affiliation: | Polytech, Area Science Park, Padriciano 99, I-34012 Trieste, Italy. miro@polytech3.area.trieste.it |
| |
Abstract: | Milk and microbial xanthine oxidases (XOs) were used for the construction of amperometric enzyme electrodes. Substrate specificity differences of these enzymes were studied. Of the two enzymes, only the microbial XO was found to oxidize theophylline, but not theobromine and caffeine. The substrate specificity of microbial XO was affected by pH, where the optimum for xanthine was 5.5, while for theophylline it was in the range from 6.5 to 8.5. The theophylline biosensor showed a low detection limit of 2 x 10(-7) M and signal linearity up to 5 x 10(-5) M. The sensitivity of the microbial XO electrode to theophylline could be selectively eliminated by immersion in alkaline phosphate solution, thus allowing for the construction of a blank electrode for differential measurements. The feasibility of this approach has been demonstrated by the determination of free (unbound) and total theophylline in blood samples. The biosensor exhibited good operational (>6 h) and shelf (>3 months) stability when trehalose was used as a stabilizer of the biocatalytic layer. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|