首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Retinal Arteriolar Morphometry Based on Full Width at Half Maximum Analysis of Spectral-Domain Optical Coherence Tomography Images
Authors:Yu Hua Tong  Tie Pei Zhu  Ze Lin Zhao  Hai Jing Zhan  Fang Zheng Jiang  Heng Li Lian
Institution:1. Quzhou People''s Hospital, Quzhou Central Hospital of Zhejiang Chinese Medical University, Quzhou, Zhejiang Province, China.; 2. Eye Center of Affiliated Second Hospital, Medical College of Zhejiang University, Hangzhou, Zhejiang Province, China.; 3. Eye Hospital of Wenzhou Medical University, Zhejiang Eye Hospital, Wenzhou, Zhejiang Province, China.; University of Melbourne, AUSTRALIA,
Abstract:

Objectives

In this study, we develop a microdensitometry method using full width at half maximum (FWHM) analysis of the retinal vascular structure in a spectral-domain optical coherence tomography (SD-OCT) image and present the application of this method in the morphometry of arteriolar changes during hypertension.

Methods

Two raters using manual and FWHM methods measured retinal vessel outer and lumen diameters in SD-OCT images. Inter-rater reproducibility was measured using coefficients of variation (CV), intraclass correlation coefficient and a Bland-Altman plot. OCT images from forty-three eyes of 43 hypertensive patients and 40 eyes of 40 controls were analyzed using an FWHM approach; wall thickness, wall cross-sectional area (WCSA) and wall to lumen ratio (WLR) were subsequently calculated.

Results

Mean difference in inter-rater agreement ranged from -2.713 to 2.658 μm when using a manual method, and ranged from -0.008 to 0.131 μm when using a FWHM approach. The inter-rater CVs were significantly less for the FWHM approach versus the manual method (P < 0.05). Compared with controls, the wall thickness, WCSA and WLR of retinal arterioles were increased in the hypertensive patients, particular in diabetic hypertensive patients.

Conclusions

The microdensitometry method using a FWHM algorithm markedly improved inter-rater reproducibility of arteriolar morphometric analysis, and SD-OCT may represent a promising noninvasive method for in vivo arteriolar morphometry.