Determination of lycopene, α-carotene and β-carotene in serum by liquid chromatography–atmospheric pressure chemical ionization mass spectrometry with selected-ion monitoring
Authors:Teruhiko Hagiwara  Tetsuko Yasuno  Keiichi Funayama  Sukeji Suzuki
Abstract:A selected-ion monitoring (SIM) determination of serum lycopene, α-carotene and β-carotene by an atmospheric pressure chemical ionization mass spectrometry (APCI–MS) was developed. A large amount of serum cholesterols disturbed the SIM determination of carotenoids by contaminating the segment of interface with the LC–MS. Therefore, separation of carotenoids from the cholesterols was performed using a mixed solution of methanol and acetonitrile (70:30) as the mobile phase on a C18 column of mightsil ODS-5 (75 mm×4.6 mm I.D.). The SIM determination was carried out by introducing only the peak portions of carotenoids and I.S. (squalene) by means of an auto switching valve. In the positive mode of APCI–MS, lycopene, α-carotene and β-carotene were monitored at m/z 537 and I.S. was monitored at m/z 411. This method was linear for all analytes in the range of 15–150 ng for lycopene, 7–70 ng for α-carotene and 25–50 ng for β-carotene. The detection limit of LC–APCI–MS-SIM for carotenoids was about 3 ng per 1 ml of serum (S/N=3). The repeatabilities, expressed as C.V.s, were 10%, 8.4% and 5.3% for lycopene, α-carotene and β-carotene, respectively. The intermediate precisions, expressed as C.V.s, were 11. 2%, 8.8% and 6.5% for lycopene, α-carotene and β-carotene, respectively.
Keywords:Lycopene  Carotene
本文献已被 ScienceDirect 等数据库收录!