A simple preservative for flow cytometric DNA analysis |
| |
Authors: | K A Alanen P J Klemi S Taimela H Joensuu |
| |
Affiliation: | Department of Pathology, University of Turku, Finland. |
| |
Abstract: | A solution containing citric acid buffered saline (CABS) and 99% ethanol (E) 1:1 was used for preserving cells for flow cytometric DNA analysis. DNA histograms obtained from fine needle biopsy aspirates and preserved in CABS+E had a similar mean coefficient of variation (CV) as was obtained from aspirates taken in CABS (3.3 vs. 3.4%) and a clearly smaller mean CV than was obtained from aspirates preserved in 50% ethanol (mean 4.8%, P less than .0001). Aspirates taken in CABS more often contained a small (less than 3,000) number of cells as compared with aspirates preserved either in CABS+E or ethanol (P less than .0001). Since preservation of cells in CABS+E allows long-term storage of samples and results in a decreased number of insufficient samples as compared with buffered saline and in an enhanced resolution as compared with 50% ethanol, CABS+E is recommended for preservation of cytological samples to be analyzed for DNA content with flow cytometry. |
| |
Keywords: | |
|
|