Simultaneous determination of citalopram and its metabolites by high-performance liquid chromatography with column switching and fluorescence detection by direct plasma injection
Authors:Eiji Matsui  Masanori Hoshino  Akiko Matsui  Akira Okahira
Abstract:High-performance liquid chromatography with a successive column-switching technique was developed for simultaneous determination of citalopram and its four metabolites in plasma. Plasma samples were injected directly, and the target compounds were purified and concentrated with an inexpensive commercial octadecyl guard column. Then, the six-port valve was switched, and the compounds retained in the column were eluted by the back-flush method using 20 mM phosphate buffer (pH 4.6)-acetonitrile (70:30, v/v) containing 0.1% diethylamine and separated with an ODS column. The compounds were assayed with a fluorescence detector at an excitation wavelength of 249 nm and an emission wavelength of 302 nm. At least 30 plasma samples could be treated with an octadecyl guard column. The limits of quantitation of this method were 2.0 ng/ml for citalopram, desmethylcitalopram, didesmethylcitalopram, citalopram propionic acid and citalopram N-oxide. This method was applied to a pharmacokinetic study in dogs and a toxicokinetic study in rats.