Accelerated liquid isolectric focusing using a direct current source that maintain constant power.
Authors:A D Strosberg  R A Jeffery  L J Freier  W J Connolly
Affiliation:1. Cardiac Unit, Massachusetts General Hospital, Boston, Massachusetts 02114 USA;2. the Department of Medicine, Harvard Medical School, Boston, Massachusetts 02114 USA;3. Skipper B Group, C. S. Draper Laboratory, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139 USA
Abstract:A power supply has been designed to improve liquid isoelectric focusing by automatically maintaining constant power applied to the column. The main advantages are (a) a reduced exposure time which is particularly important when sensitive proteins are being resolved; (b) a reduction in the loss of resolution due to diffusion or denaturation of the proteins; and (c) a considerable increase in efficiency of the use of the equipment for preparative purposes.