Precision measurements of the RSA method using a phantom model of hip prosthesis
Authors:Mäkinen Tatu J  Koort Jyri K  Mattila Kimmo T  Aro Hannu T
Institution:Departments of Surgery, University of Turku and University Central Hospital of Turku, Turku, Finland.
Abstract:Radiostereometric analysis (RSA) has become one of the recommended techniques for pre-market evaluation of new joint implant designs. In this study we evaluated the effect of repositioning of X-ray tubes and phantom model on the precision of the RSA method. In precision measurements, we utilized mean error of rigid body fitting (ME) values as an internal control for examinations. ME value characterizes relative motion among the markers within each rigid body and is conventionally used to detect loosening of a bone marker. Three experiments, each consisting of 10 double examinations, were performed. In the first experiment, the X-ray tubes and the phantom model were not repositioned between one double examination. In experiments two and three, the X-ray tubes were repositioned between one double examination. In addition, the position of the phantom model was changed in experiment three. Results showed that significant differences could be found in 2 of 12 comparisons when evaluating the translation and rotation of the prosthetic components. Repositioning procedures increased ME values mimicking deformation of rigid body segments. Thus, ME value seemed to be a more sensitive parameter than migration values in this study design. These results confirmed the importance of standardized radiographic technique and accurate patient positioning for RSA measurements. Standardization and calibration procedures should be performed with phantom models in order to avoid unnecessary radiation dose of the patients. The present model gives the means to establish and to follow the intra-laboratory precision of the RSA method. The model is easily applicable in any research unit and allows the comparison of the precision values in different laboratories of multi-center trials.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号