首页 | 本学科首页   官方微博 | 高级检索  
     


Separation of D- and L-amino acids by ion exchange column chromatography in the form of alanyl dipeptides
Authors:J. Csapó  I. Tóth-Pósfai  ZS. Csapó-Kiss
Affiliation:(1) Faculty of Animal Science Kaposvár, Pannon Agricultural University, Dénesmajor 2, H-7401 Kaposvár, Hungary
Abstract:Summary A method of ion exchange column chromatography was developed for the determination of D- and L-amino acids in the form of diastereomeric dipeptide. First the protein containing samples were hydrolyzed with 6 molar hydrochloric acid, then the single amino acids were separated in an LKB automated amino acid analyzer with the LKB fraction collector. Following lyophilization, the single amino acids were transformed into alanyl dipeptides with tertiary-butyloxycarbonil-L-alanine-N-hydroxy-succinimide (t-BOC-L-Ala-ONSu) active ester. The alanyl dipeptides were easily separated from one another and the initial amino acids. Determination of the D- and L-amino acids in this form is relatively accurate and reproducible but takes some time (33–38 min). Accuracy of the determination is satisfactory. The coefficient of variation amounts to 3–5%. The use of the method is suggested to laboratories having an amino acid analyzer and wish to determine D-and L-amino acids in synthetic-amino acids complements, peptides or natural materials.
Keywords:Amino acids  D- and L-amino acid determination  Diastereomeric dipeptides  Alanyl dipeptides  Ion exchange column chromatography
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏