首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Determination of 9α,11β-prostaglandin F2 in human urine and plasma by gas chromatography—mass spectrometry
Authors:T Obata  T Nagakura  M Kammuri  T Masaki  K Maekawa  K Yamashita
Abstract:Sensitive and specific assay methods for 9α,11β-prostaglandin F2 (9α,11β-PGF2) by gas chromatography—mass spectrometry with electron impact ionization are described. The mass spectrometric assay for 9α,11β-PGF2 was based on the use of the methyl ester—dimethylisopropylsilyl ether derivative, and pentadeuterated PGF as a convenient internal standard. The calibration graph for 9α,11β-PGF2 was linear from 5 pg to 100 ng for both the standard and spiked biological samples. The limit of detection was 50 pg/ml for urine and 25 pg/ml for plasma (signal-to-noise RATIO = 2.3). The method was applied to the determination of 9α,11β-PGF2 in urine and plasma samples from patients with bronchial asthma.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号