Detect of Escherichia coli O157:H7 in ground beef samples using piezoelectric excited millimeter-sized cantilever (PEMC) sensors |
| |
Authors: | Campbell Gossett A Uknalis Joe Tu Shu-I Mutharasan Raj |
| |
Affiliation: | Department of Chemical and Biological Engineering, Drexel University, Philadelphia, PA 19104, United States. |
| |
Abstract: | Piezoelectric-excited millimeter-sized cantilever (PEMC) sensors consisting of a piezoelectric and a borosilicate glass layer with a sensing area of 4 mm2 were fabricated. An antibody specific to Escherichia coli (anti-E. coli) O157:H7 was immobilized on PEMC sensors, and exposed to samples containing E. coli O157:H7 (EC) prepared in various matrices: (1) broth, broth plus raw ground beef, and broth plus sterile ground beef without inoculation of E. coli O157:H7 served as controls, (2) 100 mL of broth inoculated with 25 EC cells, (3) 100 mL of broth containing 25 g of raw ground beef and (4) 100 mL of broth with 25 g of sterile ground beef inoculated with 25 EC cells. The total resonant frequency change obtained for the broth plus EC samples were 16+/-2 Hz (n=2), 30 Hz (n=1), and 54+/-2 Hz (n=2) corresponding to 2, 4, and 6h growth at 37 degrees C, respectively. The response to the broth plus 25 g of sterile ground beef plus EC cells were 21+/-2 Hz (n=2), 37 Hz (n=1), and 70+/-2 Hz (n=2) corresponding to 2, 4, and 6 h, respectively. In all cases, the three different control samples yielded a frequency change of 0+/-2 Hz (n=6). The E. coli O157:H7 concentration in each broth and beef samples was determined by both plating and by pathogen modeling program. The results indicate that the PEMC sensor detects E. coli O157:H7 reliably at 50-100 cells/mL with a 3 mL sample. |
| |