首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Detection and quantification of Legionella pneumophila in water samples using competitive PCR
Authors:Declerck Priscilla  Behets Jonas  Lammertyn Elke  Lebeau Ilya  Anné Jozef  Ollevier Frans
Institution:Laboratory of Aquatic Ecology, Zoological Institute, Katholieke Universiteit Leuven, Belgium. Priscilla.declerck@bio.kuleuven.be
Abstract:The presence of high levels of Legionella pneumophila in man-made aquatic systems correlates with the incidence of nosocomial Legionnaires' disease. This requires a rapid, reliable, and sensitive quantification of L. pneumophila concentrations in suspected water systems. In this research, a homologous competitor was developed and evaluated in a L. pneumophila competitive polymerase chain reaction (cPCR) to quantify this human pathogen in a quick, cost-effective, and reliable way. Accuracy of cPCR was evaluated by analyzing cooling tower and tap water samples spiked with known concentrations of L. pneumophila bacteria, in parallel with the standard culture method. Legionella pneumophila amounts detected and calculated from cPCR and culture correlated very well: r = 0.998, P = 0.002 for tap water and r = 0.990, P = 0.009 for cooling tower water. Nevertheless, for both kinds of water samples, mean numbers of L. pneumophila calculated from cPCR results were always higher than those obtained by culture. This study makes it clear that the rapid, sensitive, and cost-effective L. pneumophila cPCR is a promising alternative to the standard time-consuming culture method and expensive real-time PCR to enumerate L. pneumophila bacteria in environmental water samples.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号