Critical assessment of the steady-state Na2SO3 feeding method for kla measurement in fermentors
Authors:Linek V  Benes P  Sinkule J
Institution:Institute of Chemical Technology, Department of Chemical Engineering, 16628 Prague 6 Czechoslovakia.
Abstract:Important aspects of k(l)a measurement in agitated aerated vessels are briefly characterized from the standpoint of reliability of the measured data. It seems that most of the k(l)a data, based on a number of variants of the steady-state and dynamic methods in noncoalescent liquids, do not have a clear physical meaning, because they are affected by the differences between the actual driving force and the driving force assumed by the model used for its evaluation. A reliability test is given for the Na(2)SO(3) feeding steady-state method (FSM), by comparing the results of air and pure oxygen absorption in a noncoalescent liquid (0.5M Na(2)SO(4) solution) with the results obtained by the independent pressure step dynamic method (RDM). The RDM is one of a few variants of the dynamic method which gives correct k(l)a data unaffected by nonideal mixing of the gas phase in the reactor. It was found that the FSM yields correct k(l)a values only when pure oxygen is used for absorption. When air is absorbed, the FSM gives k(l)a values in the region of k(l)a > 0.1 s(-1) substantially (to 55%) lower than those for pure oxygen absorption.