Rapid and sensitive LC–MS/MS assay for the quantitation of 20(S)-protopanaxadiol in human plasma
Authors:Dan Zhang  Yingwu Wang  Jiangbin Han  Weisong Yu  Lili Deng  J Paul Fawcett  Zeyuan Liu  Jingkai Gu
Institution:1. Research Center for Drug Metabolism, Jilin University, Changchun 130023, PR China;2. School of Pharmacy, University of Otago, P.O. Box 913, Dunedin, New Zealand;3. Department of Pharmacology, Affiliated Hospital of the Academy of Military Medical Sciences, Beijing 100071, PR China
Abstract:This paper describes a rapid and sensitive method for the quantitation of 20(S)-protopanaxadiol (PPD) in human plasma based on high-performance liquid chromatography–tandem mass spectrometry (LC–MS/MS). The analyte and internal standard (I.S.), ginsenoside Rh2, were extracted from plasma by liquid–liquid extraction and separated on a Zorbax extend C18 analytical column using methanol–acetonitrile-10 mM ammonium acetate (47.5:47.5:5, v/v/v) as mobile phase. Detection was by tandem mass spectrometry using electrospray ionization in the positive ion mode and multiple reaction monitoring (MRM). The assay was linear over the concentration range 0.1–100.0 ng/ml with a limit of detection of 0.05 ng/ml. The method was successfully applied to a clinical pharmacokinetic study in healthy volunteers after a single oral administration of a PPD 25 mg capsule.