Piezoelectric detection of bilirubin based on bilirubin-imprinted titania film electrode |
| |
Authors: | Yang Zhengpeng Yan Jinlong Zhang Chunjing |
| |
Affiliation: | Institute of Materials Science and Engineering, Henan Polytechnic University, Jiaozuo 454000, China. zhengpengyang@yahoo.com.cn |
| |
Abstract: | A novel quartz crystal microbalance (QCM) sensor with a high selectivity and sensitivity has been developed for bilirubin determination, based on the modification of bilirubin-imprinted titania film onto a quartz crystal by molecular imprinting and surface sol-gel techniques. The performance of the developed bilirubin biosensor was evaluated and the results indicated that a sensitive bilirubin biosensor could be fabricated. The obtained bilirubin biosensor presents high-selectivity monitoring of bilirubin, better reproducibility, shorter response time (30 min), wider linear range (0.1-50 μM), and lower detection limit (0.05 μM). The analytical application of the bilirubin biosensor confirms the feasibility of bilirubin determination in serum sample. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|