首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Q-band electron paramagnetic resonance dosimetry in tooth enamel: biopsy procedure and determination of dose detection limit
Authors:Alex Romanyukha  François Trompier  Ricardo A Reyes
Institution:1. Naval Dosimetry Center, Bethesda, MD, USA
2. Institut de Radioprotection et de S?reté Nucléaire, Fontenay-aux-Roses, France
3. Uniformed Services University of the Health Sciences, Bethesda, MD, USA
Abstract:High-frequency Q-band (37 GHz) electron paramagnetic resonance (EPR) dosimetry allows to perform fast (i.e., measurement time <15 min) dose measurements using samples obtained from tooth enamel mini-biopsy procedures. We developed and tested a new procedure for taking tooth enamel biopsy for such dose measurements. Recent experience with EPR dose measurements in Q-band using mini-probes of tooth enamel has demonstrated that a small amount of tooth enamel (2–10 mg) can be quickly obtained from victims of a radiation accident. Accurate dose assessments can further be carried out in a very short time to provide important information for medical treatment. Here, the Q-band EPR dose detection limit for 5 and 10 mg samples is estimated to be 367 and 248 mGy, respectively. These values are comparable to the critical parameters determined for conventional X-band EPR in tooth enamel.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏