Abstract:A high-performance liquid chromatographic method with diode array detection (HPLC-DAD) has been developed for the simultaneous separation and quantification of imipramine, amitriptyline, maprotiline, fluoxetine, clomipramine and their respective metabolites using a 500-μl plasma sample and clovoxamine as the internal standard. The substances were eluted on a Symmetry C18 5-μm column (250×4.6 mm, I.D.). Full UV spectra from 200 to 450 nm were recorded on-line during the entire analysis and were automatically compared to spectra stored in a library. The quantification was performed at 226, 254, and 400 nm. Peak height ratios were linear over a concentration range of 10–3000 ng/ml: the correlation coefficient (r) was better than 0.998 for all substances at each wavelength. Acceptable coefficients of variation are demonstrated for both within-run and day-to-day assays. The method is simple, highly specific and currently being used for drug monitoring and toxicological studies in children and adult patients.