Dual wavelength/stopped flow spectrophotometry: computer acquisition and analysis.
Authors:J A Peterson  D M Mock
Affiliation:1. Department of Biochemistry, The University of Texas Health Science Center, Dallas, Texas, USA;2. Southwestern Medical School 5323 Hines Boulevard, Dallas, Texas 75235 USA
Abstract:The adaptation of a commercially available dual wavelength/stopped flow spectrophotometer for use with turbid samples is described. A minicomputer is used to collect and analyze the data, thereby facilitating these experiments. The stopped flow/computer combination has a dead time in the single wavelength mode of 3.5 msec. In the dual wavelength mode, accurate determinations can be made of the time course of reactions that have a t12 of 50 msec or longer. The application of this stopped flow spectrophotometer to the measurement of cytochrome P-450 reductase activity in rat liver microsomes is described.