An automatic method for determination of urinary 3-methylhistidine: normal values. |
| |
Authors: | Monika Neuhäuser Peter Fürst |
| |
Affiliation: | Metabolic Research Laboratory, Sit Eriks sjukhus, Stockholm, Sweden |
| |
Abstract: | A system for automatic analysis of urinary 3-methylhistidine is described, applying ion-exchange chromatography and using an automatic sample injector, a motoric selector valve, and a diode programmer, which controls the analytical system. The method permits a sampling rate of 22 samples/day. 3-Methylhistidine was completely separated from histidine in 37 min whereas 1-methylhistidine was eluted together with ammonia. The 3-methylhistidine concentration was linear up to 150 nmol/ml and no appreciable sample interaction was found at automatic sequential runs. The error, in a single determination based on duplicate samples, was 4.61% and, in duplicated determinations, 3.26%. The mean urinary 3-methylhistidine output was 299.4 ± 23.8 μmol/day in 12 healthy females and 545.5 ± 35.2 μmol/day in 12 healthy males. The 3-methylhistidine excretion was significantly higher in males than in females, when expressed as the absolute daily output or as the estimated ratio to body weight, body surface area, or creatinine. |
| |
Keywords: | To whom correspondence should be addressed. |
本文献已被 ScienceDirect 等数据库收录! |
|