首页 | 本学科首页   官方微博 | 高级检索  
     


Determination by high-performance liquid chromatography of hydroxyurea in human plasma
Authors:J. Havard
Affiliation:

Department of Biochemistry, Royal North Shore Hospital, Pacific Highway, St. Leonards, N.S.W. 2065 Australia

Department of Clinical Oncology, Royal Prince Alfred Hospital, Camperdown, N.S.W. 2050 Australia

Department of Biochemistry, Royal North Shore Hospital, Pacific Highway, St. Leonards, N.S.W. 2065 Australia

Abstract:An assay using reversed-phase high-performance liquid chromatography with electrochemical detection was developed to measure hydroxyurea in plasma at concentrations suitable for pharmacokinetic studies. The sample preparation is simple, the analysis rapid and assays can be batched. The between-run precision is excellent (coefficient of VARIATION = 2.8–4.5%) and the limit of detection is 0.02 mmol/l. Preliminary studies have shown that the method is suitable for pharmacokinetic studies.
Keywords: