Immunoassay-based determination of phenobarbital using size-exclusion chromatography |
| |
Authors: | Andrea Bereczki, Viola Horv th,George Horvai |
| |
Affiliation: | a Institute of General and Analytical Chemistry, Budapest University of Technology and Economics, Gellért tér 4, H-1111 Budapest, Hungary;b Division of Chemical Information Technology, Budapest University of Technology and Economics, Gellért tér 4, H-1111 Budapest, Hungary |
| |
Abstract: | The measurement of the anti-epileptic drug phenobarbital from serum samples combining immunoassay and size-exclusion chromatography is presented. The immunoreaction is based on the competitive binding of the analyte (unlabelled phenobarbital) and the fluorescent-labelled phenobarbital to anti-phenobarbital antibodies. Mixing of the reagents and the immunoreaction takes place in a flow system. The products are separated on-line on a short gel chromatographic column and the fluorescence intensity of the marker is measured. The calibration curve shows good linearity in the range 5–80 μg/ml, corresponding to therapeutically relevant serum levels. Intra-day precision values are between 7.32 and 9.48%; the accuracy is between 0.97 and 9.43%. Inter-day precision and accuracy measured on 6 different days fall between 5.38 and 10.05% and −8.27 and −4.97%, respectively. The results obtained with the proposed method show a good correlation with those of other methods (radioimmunoassay and fluorescence polarisation immunoassay) already established in clinical laboratories. |
| |
Keywords: | Phenobarbital Fluorescent immunoassay Size-exclusion chromatography |
本文献已被 ScienceDirect 等数据库收录! |
|