1. Division of Molecular Pharmacology, National Institute for Medical Research, Mill Hill, London, NW7 1AA, United Kingdom;2. Department of Chemistry, University of South Carolina, Columbia, South Carolina 29208 USA
Abstract:
It is shown that absorption of the excitation light can lead to substantial systematic errors in fluorescence measurements of equilibrium constants for formation of protein-ligand complexes. The assumptions about the optical arrangement of the fluorescence spectrometer involved in the calculation of the correction of this absorption are discussed. A general semiempirical correction procedure which can be used for (calculated) absorbance values as high as 5 is described. The importance of choosing the excitation wavelength so as to minimize the necessity for these corrections is emphasized.