首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Comparison of arrhythmia detection by conventional Holter and a novel ambulatory ECG system using patch and Android App,over 24 h period
Institution:Department of Cardiology, Government Medical College, Thrissur, Kerala, 680596, India
Abstract:BackgroundAmbulatory electrocardiogram (AECG) is done for evaluation of arrhythmia. Commonly used AECG system is 24 h Holter. Patch based second generation AECG monitoring devices, which can record for longer periods, are now available.ObjectiveAndroid App based WebCardio using WiPatch is a new AECG system which records ECG in two leads for 72 h. Our study compared the arrhythmia detection by WebCardio and conventional Holter by simultaneously connecting both for 24 h in patients having indication for AECG.MethodsThe AECG of patients who had simultaneous recording with WebCardio and conventional Holter, in the department of Cardiology, Medical College, Thrissur were evaluated. Ability to detect any of the 6 arrhythmias :1) atrial fibrillation (AF), 2) atrioventricular (AV) block, 3) sinus pause of ≥3 s (SP), 4) supraventricular tachycardia (SVT), 5) premature ventricular complex (PVC) and 6) ventricular tachycardia (VT)/ventricular fibrillation (VF) was compared. Detection of each arrhythmia was also compared.Results141 patients had simultaneous recordings by both systems of AECG. The WebCardio picked up at least one of the 6 arrhythmias; AF, AV block, SP, SVT, PVC or VT/VF in 98 cases compared to 88 in the Holter (McNemars test, two tail P = 0.006). In eleven cases WebCardio detected an arrhythmia where Holter could not. In one case Holter identified an arrhythmia and WebCardio could not. Individual arrhythmias; AF, SP, SVT and VT/VF were detected equally by both systems. AV block (23 Vs 18, p = 0.0625) and PVCs (83 Vs 74, p = 0.0636) were detected in more number of cases in WebCardio. In the five cases where WebCardio alone identified AV block, four had poor quality of P wave in the Holter.ConclusionArrhythmia was picked up in more number of patients by the WebCardio compared to Holter. This was due to higher pickup of AV block and PVCs by WebCardio. Difference in AV block identification was due to better quality of P in WebCardio. WebCardio is a good alternative to Holter for AECG.
Keywords:Holter  Ambulatory ECG  Patch  Arrhythmia
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号