首页 | 本学科首页   官方微博 | 高级检索  
     


Quantification of nisin in flow-injection immunoassay systems
Authors:Nandakumar R  Nandakumar M P  Mattiasson B
Affiliation:Department of Biotechnology, Center for Chemistry and Chemical Engineering, Lund University, Sweden.
Abstract:A monoclonal-antibody-based, sequential competitive-flow-injection immunoassay system in expanded-bed mode has been developed for the determination of nisin. The system allows the determination of nisin in the presence of suspended particles without any significant interference, illustrating its potential for on-line monitoring of fermentation processes or the analysis of food matrices. The dose response range of the system when operated in expanded-bed mode was 6-90 microM. The detection limit under packed-bed conditions was 3 microM. The results correlated well with the results from conventional ELISA in the analysis of samples of processed cheese. When milk samples, fermentation samples and buffer were spiked with nisin, the mean recoveries were 86% for milk samples, 96% for fermentation samples and 98% for buffer solution.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏