The use of isochronal reference standards in phase and modulation fluorescence lifetime measurements
Authors:David A. Barrow  Barrty R. Lentz
Affiliation:Department of Biochemistry and Nutrition, University of North Carolina, Chapel Hill, NC 27514, U.S.A.
Abstract:Errors in phase and modulation lifetime measurements observed with the only commercially available instrument are readily apparent when the Debye-Sears modulation tank is not perfectly tuned. Unfortunately, we have found that exact tuning was extremely difficult to achieve and maintain. We report that these errors could be reduced by using single-lifetime (homogeneous) reference standards whose fluorescence lifetime approximated that of the unknown sample (isochronal standards). A number of useful standards are suggested. In the proposed method, the phase shift and relative modulation of the sample emission are measured using the isochronal standard as a reference to determine the effective characteristics of the sinusoidal excitation. The importance of the improvement in accuracy accomplished by the proposed methods is illustrated by the accurate resolution of fluorescence lifetime heterogeneity for two known heterogeneous samples.
Keywords:phase and modulation fluorescence lifetimes  lifetime heterogeneity  DPH  fluorescence standards  DPH  1,6-diphenyl-1,3,5-hexatriene  POPOP  9-AC  9-anthracenecarbonitrile  DENS  2-diethylaminonaphthalene-5-sulfonic acid