An improved apparatus for calibration of warburg manometers |
| |
Affiliation: | 1. College of mechanical and electronic engineering, Shandong University of Science and Technology, Qingdao, Shandong 266590, China;3. Shanghai Key Laboratory of Advanced High-temperature Materials and Precision Forming, Shanghai Jiao Tong University, 800 Dongchuan Road, Shanghai 200240, PR China;4. Shandong Derui Anti-Corrosion Materials Co., Ltd, Zibo, Shandong 256100, China;5. Affiliated Hospital of Youjiang Medical University for Nationalities, Baise, Guangxi 533000, China |
| |
Abstract: | - 1.1. By the use of a new piece of apparatus, a procedure is described that allows more accurate determination of gas volume in manometers than has been previously possible.
- 2.2. The standard error of the mean of ten readings of gas volume for a single manometer was found to be ±7 × 10−5 ml.
- 3.3. It is possible to obtain repeat readings rapidly and with very little spillage of mercury.
- 4.4. This method also allows easy removal of any air trapped by the rising mercury.
|
| |
Keywords: | |