A reliability and validity study for Scolioscan: a radiation-free scoliosis assessment system using 3D ultrasound imaging
Authors:Email author" target="_blank">Yong-Ping?ZhengEmail author  Timothy?Tin-Yan?Lee  Kelly?Ka-Lee?Lai  Benjamin?Hon-Kei?Yip  Guang-Quan?Zhou  Wei-Wei?Jiang  James?Chung-Wai?Cheung  Man-Sang?Wong  Bobby?King-Wah?Ng  Jack?Chun-Yiu?Cheng  Tsz-Ping?Lam
Institution:1.Interdisciplinary Division of Biomedical Engineering,The Hong Kong Polytechnic University,Hong Kong,People’s Republic of China;2.School of Public Health and Primary Care,The Chinese University of Hong Kong,Hong Kong,People’s Republic of China;3.Department of Orthopaedics and Traumatology,The Chinese University of Hong Kong,Hong Kong,People’s Republic of China
Abstract:

Background

Radiographic evaluation for patients with scoliosis using Cobb method is the current gold standard, but radiography has radiation hazards. Several groups have recently demonstrated the feasibility of using 3D ultrasound for the evaluation of scoliosis. Ultrasound imaging is radiation-free, comparatively more accessible, and inexpensive. However, a reliable and valid 3D ultrasound system ready for clinical scoliosis assessment has not yet been reported. Scolioscan is a newly developed system targeted for scoliosis assessment in clinics by using coronal images of spine generated by a 3D ultrasound volume projection imaging method. The aim of this study is to test the reliability of spine deformity measurement of Scolioscan and its validity compared to the gold standard Cobb angle measurements from radiography in adolescent idiopathic scoliosis (AIS) patients.

Methods

Prospective study divided into two stages: 1) Investigation of intra- and inter- reliability between two operators for acquiring images using Scolioscan and among three raters for measuring spinal curves from those images; 2) Correlation between the Cobb angle obtained from radiography by a medical doctor and the spine curve angle obtained using Scolioscan (Scolioscan angle). The raters for ultrasound images and the doctors for evaluating radiographic images were mutually blinded. The two stages of tests involved 20 (80 % females, total of 26 angles, age of 16.4?±?2.7 years, and Cobb angle of 27.6?±?11.8°) and 49 (69 % female, 73 angles, 15.8?±?2.7 years and 24.8?±?9.7°) AIS patients, respectively. Intra-class correlation coefficients (ICC) and Bland-Altman plots and root-mean-square differences (RMS) were employed to determine correlations, which interpreted based on defined criteria.

Results

We demonstrated a very good intra-rater and intra-operator reliability for Scolioscan angle measurement with ICC larger than 0.94 and 0.88, respectively. Very good inter-rater and inter-operator reliability was also demonstrated, with both ICC larger than 0.87. For the thoracic deformity measurement, the RMS were 2.5 and 3.3° in the intra- and inter-operator tests, and 1.5 and 3.6° in the intra- and inter-rater tests, respectively. The RMS differences were 3.1, 3.1, 1.6, 3.7° in the intra- and inter-operator and intra- and inter-rater tests, respectively, for the lumbar angle measurement. Moderate to strong correlations (R2?>?0.72) were observed between the Scolioscan angles and Cobb angles for both the thoracic and lumbar regions. It was noted that the Scolioscan angle slightly underestimated the spinal deformity in comparison with Cobb angle, and an overall regression equation y?=?1.1797x (R2?=?0.76) could be used to translate the Scolioscan angle (x) to Cobb angle (y) for this group of patients. The RMS difference between Scolioscan angle and Cobb angle was 4.7 and 6.2°, with and without the correlation using the overall regression equation.

Conclusions

We showed that Scolioscan is reliable for measuring coronal deformity for patients with AIS and appears promising in screening large numbers of patients, for progress monitoring, and evaluation of treatment outcomes. Due to it being radiation-free and relatively low-cost, Scolioscan has potential to be widely implemented and may contribute to reducing radiation dose during serial monitoring.
Keywords: