Thin-film antimony-antimony-oxide enzyme electrode for penicillin determination |
| |
Authors: | Flanagan M T Carroll N J |
| |
Affiliation: | Department of Electronic and Electrical Engineering, University College London, Torrington Place, London, WC1E 7JE, England. |
| |
Abstract: | A potentiometric penicillinase electrode is reported in which the base pH transducer is a thin-film anti-mony-antimony-oxide electrode deposited by vacuum evaporation. Several enzyme immobilization procedures have been examined and a crosslinked protein film found to be the most appropriate to this type of sensor. The use of an adjacent antimony-antimony-oxide track as a pseudoreference electrode was successfully demonstrated. The overall response was shown to be independent of the stirring rate above 100 rpm, but the kinetics of the response were found to depend markedly on the stirring rate. The intrinsic linear response range was 3 x 10(-4)Mto 7 x 10(-3)M penicillin G. Linearizing transforms that extend the useful range were examined. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|