Abstract: | Measuring turbulent shear stresses is of major importance in artificial heart valve evaluation. Bi- and unidirectional fluid velocity measurements enable calculation of Reynolds shear stress () and Reynolds normal stress (). τ is important due to the relation to hemolysis and thrombus formation, but σ is the only obtainable parameter in vivo. Therefore, determination of a correlation factor between τ and σ is pertinent.In a pulsatile flow model, laser Doppler (LDA) and hot-film (HFA) anemometry were used for simultaneous bi- and unidirectional fluid velocity measurements downstream of a Hall Kaster and a Hancock Porcine aortic valve. Velocities were registered in two flow field locations and at four cardiac outputs. The velocity signals were subjected to analog signal processing prior to digital turbulence analysis, as a basis for calculation of τ and σ.A correlation factor of 0.5 with a correlation coefficient of 0.97 was found between the maximum Reynolds shear stress and Reynolds normal stress, implying . In vitro estimation of turbulent shear stresses downstream of artificial aortic valves, based on the axial velocity component alone, seems possible. |