首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Biomechanical Analysis of a Newly Developed Shape Memory Alloy Hook in a Transforaminal Lumbar Interbody Fusion (TLIF) In Vitro Model
Authors:Xi Wang  Jing Xu  Yuexing Zhu  Jiukun Li  Si Zhou  Shunliang Tian  Yucheng Xiang  Xingmo Liu  Ying Zheng  Tao Pan
Institution:1. Department of Orthopaedic Surgery, The Sixth Affiliated Hospital of Sun Yat-sen University, Guangzhou, PR China.; 2. Department of Anatomy, Guilin Medical College, Guilin, PR China.; 3. Department of Nutrition, Guangdong General Hospital, Guangzhou, PR China.; University of Michigan, United States of America,
Abstract:

Objective

The objective of this biomechanical study was to evaluate the stability provided by a newly developed shape memory alloy hook (SMAH) in a cadaveric transforaminal lumbar interbody fusion (TLIF) model.

Methods

Six human cadaveric spines (L1-S2) were tested in an in vitro flexibility experiment by applying pure moments of ±8 Nm in flexion/extension, left/right lateral bending, and left/right axial rotation. After intact testing, a TLIF was performed at L4-5. Each specimen was tested for the following constructs: unilateral SMAH (USMAH); bilateral SMAH (BSMAH); unilateral pedicle screws and rods (UPS); and bilateral pedicle screws and rods (BPS). The L3–L4, L4–L5, and L5-S1 range of motion (ROM) were recorded by a Motion Analysis System.

Results

Compared to the other constructs, the BPS provided the most stability. The UPS significantly reduced the ROM in extension/flexion and lateral bending; the BSMAH significantly reduced the ROM in extension/flexion, lateral bending, and axial rotation; and the USMAH significantly reduced the ROM in flexion and left lateral bending compared with the intact spine (p<0.05). The USMAH slightly reduced the ROM in extension, right lateral bending and axial rotation (p>0.05). Stability provided by the USMAH compared with the UPS was not significantly different. ROMs of adjacent segments increased in all fixed constructs (p>0.05).

Conclusions

Bilateral SMAH fixation can achieve immediate stability after L4–5 TLIF in vitro. Further studies are required to determine whether the SMAH can achieve fusion in vivo and alleviate adjacent segment degeneration.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏