首页 | 本学科首页   官方微博 | 高级检索  
     


Corticosteroid analysis in biological fluids by high-performance liquid chromatography
Authors:James Q. Rose  William J. Jusko
Affiliation:Clinical Pharmacokinetics Laboratory, Department of Pharmaceutics, School of Pharmacy, State University of New York at Buffalo and Millard Fillmore Hospital, 3 Gates Circle, Buffalo, N.Y. 14209 U.S.A.
Abstract:A sensitive, specific, and reproducible high-performance liquid chromatographic assay for the simultaneous determination of prednisone, prednisolone and cortisol in biological fluids was developed with dexamethasone as the internal standard. Samples are extracted with methylene chloride, washed with sodium hydroxide and then water, and chromatographed on a microparticulate silica gel column with UV detection at 254 nm. Sensitivity was greater than 15 ng for all four steroids. Specificity was supported by use of dual wavelength UV detection and/or radioimmunoassay. The assay has been applied in pharmacokinetic studies and a typical plasma concentration—time profile for the three steroids is presented for one subject who received 50 mg of prednisone.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号