Quantification of the effect of electrical and thermal parameters on radiofrequency ablation for concentric tumour model of different sizes
Institution:1. Laboratoire Quartz EA 7393, École Supérieure d’Ingénieurs en Génie Électrique, Productique et Management Industriel, Cergy Pontoise Cedex, France;2. Laboratoire de Recherche en Eco-innovation Industrielle et Énergétique, École Supérieure d’Ingénieurs en Génie Électrique, Productique et Management Industriel, Cergy Pontoise Cedex, France;1. Physics Department, University of Athens, 15784 Panepistimioupolis, Athens, Greece;2. CEA, LIST, Diamond Sensors Lab., F-91191 Gif-sur-Yvette, France
Abstract:Radiofrequency ablation (RFA) has been increasingly used in treating cancer for multitude of situations in various tissue types. To perform the therapy safely and reliably, the effect of critical parameters needs to be known beforehand. Temperature plays an important role in the outcome of the therapy and any uncertainties in temperature assessment can be lethal. This study presents the RFA case of fixed tip temperature where we’ve analysed the effect of electrical conductivity, thermal conductivity and blood perfusion rate of the tumour and surrounding normal tissue on the radiofrequency ablation. Ablation volume was chosen as the characteristic to be optimised and temperature control was achieved via PID controller. The effect of all 6 parameters each having 3 levels was quantified with minimum number of experiments harnessing the fractional factorial characteristic of Taguchi’s orthogonal arrays. It was observed that as the blood perfusion increases the ablation volume decreases. Increasing electrical conductivity of the tumour results in increase of ablation volume whereas increase in normal tissue conductivity tends to decrease the ablation volume and vice versa. Likewise, increasing thermal conductivity of the tumour results in enhanced ablation volume whereas an increase in thermal conductivity of the surrounding normal tissue has a debilitating effect on the ablation volume and vice versa. With increase in the size of the tumour (i.e., 2–3 cm) the effect of each parameter is not linear. The parameter effect varies with change in size of the tumour that is manifested by the different gradient observed in ablation volume. Most important is the relative insensitivity of ablation volume to blood perfusion rate for smaller tumour size (2 cm) that is also in accordance with the previous results presented in literature. These findings will provide initial insight for safe, reliable and improved treatment planning perceptively.
Keywords:Radiofrequency ablation  PID controller  Taguchi’s orthogonal arrays
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号