In-vivo dose measurements with MOSFET dosimeters during MV portal imaging |
| |
Authors: | Sathish Kumar Rabi Raja Singh Henry Finlay Godson Retna Ponmalar Paul Ravindran Sunil Dutt Sharma Subhashini John |
| |
Affiliation: | 1.Department of Radiation Oncology, Christian Medical College, Vellore, India; 2.Radiological Physics and Advisory Division, Bhabha Atomic Research Center, Mumbai, India |
| |
Abstract: | BackgroundThe purpose of this study was to investigate the feasibility of MOSFET dosimeter in measuring eye dose during 2D MV portal imaging for setup verification in radiotherapy.Materials and methodsThe in-vivo dose measurements were performed by placing the dosimeters over the eyes of 30 brain patients during the acquisition of portal images in linear accelerator by delivering 1 MU with the field sizes of 10 × 10 cm2 and 15 × 15 cm2.ResultsThe mean doses received by the left and right eyes of 10 out of 30 patients when both eyes were completely inside the anterior portal field were found to be 2.56 ± 0.2 cGy and 2.75 ± 0.2, respectively. Similarly, for next 10 patients out of the same 30 patients the mean doses to left and right eyes when both eyes were completely out of the anterior portal fields were found to be 0.13 ± 0.02 cGy and 0.17 ± 0.02 cGy, respectively. The mean doses to ipsilateral and contralateral eye for the last 10 patients when one eye was inside the anterior portal field were found to be 3.28 ± 0.2 cGy and 0.36 ± 0.1 cGy, respectively.ConclusionThe promising results obtained during 2D MV portal imaging using MOSFET have shown that this dosimeter is well suitable for assessing low doses during imaging thereby enabling to optimize the imaging procedure using the dosimetric data obtained. In addition, the documentation of the dose received by the patient during imaging procedure is possible with the help of an in-built software in conjunction with the MOSFET reader module. |
| |
Keywords: | MV portal imaging setup verification in vivo dosimetry MOSFET |
|
|