Quantification of raltegravir (MK0518) in human plasma by high-performance liquid chromatography with photodiode array detection
Authors:Lauriane Goldwirt  Aurélie Barrail-Tran  Maria Da Cruz  Anne-Marie Taburet  Valérie Furlan
Institution:Pharmacie Clinique, Laboratoire de Toxicologie, CHU de Bicêtre AP-HP, 78 avenue du Général Leclerc, 94270 Le Kremlin-Bicêtre, France
Abstract:A precise and accurate high-performance liquid chromatography (HPLC) method with photodiode array detection has been developed and validated for raltegravir, a human immunodeficiency virus integrase strand transfer inhibitor (HIV-1 INSTI). Plasma (300 μL) was extracted with dichloromethane/hexane 50:50 (v/v) after addition of the internal standard, 6,7-dimethyl-2,3-di(2-pyridyl) quinoxaline. The compounds were separated using a dC18 column and detected with ultraviolet detection at 320 nm. The limit of quantification was 10 ng/mL for raltegravir. The method was linear and validated over a concentration range of 0–10,000 ng/mL. The intra-day precision ranged from 3.1 to 12.3%, while the intra-day accuracy ranged from ?15.0 to ?0.5%, the inter-day precision and accuracy were less than 7%. The mean recovery was 76.8%. Application to clinical samples taken from patients treated with raltegravir indicated that the method is suitable for measuring plasma concentrations of raltegravir in pharmacokinetic studies of clinical trials.