On the isolation of polychlorinated dibenzo-p-dioxins and furans from serum samples using immunoaffinity chromatography prior to high-resolution gas chromatography–mass spectrometry
Authors:Janice K Huwe  Weilin L Shelver  Larry Stanker  Donald G Patterson  Jr  Wayman E Turner
Abstract:Immunoaffinity chromatography (IAC) for the purification of polychlorinated dibenzo-p-dioxins and furans (PCDD/Fs) from biological samples was explored as a means to simplify the cleanup procedure and thereby decrease the time and cost of dioxin analysis. A monoclonal antibody (DD3) was used to produce IAC columns and to isolate the PCDD/Fs from serum. Native and 13C-labeled PCDD/Fs were spiked at the ppq to ppt range into serum. Quantitation of the PCDD/Fs was performed by a standard dioxin analytical method, i.e. high-resolution gas chromatography–mass spectrometry (GC–MS), which was easily compatible with IAC. Five of the most toxic PCDD/Fs consistently showed acceptable recoveries (>25%) and were reliably quantitated. The congeners specifically recovered by this method represent almost 80% of the toxic equivalency of dioxins and furans present in the serum samples. Dioxin-like polychlorinated biphenyls (PCBs) were not recognized by this antibody column. Compared to conventional dioxin cleanup methods, IAC decreased solvent usage by 1.5 l/sample and took only 2 h to process a sample for analysis.
Keywords:Polychlorinated dibenzo-p-dioxins  Furans