Determination of phloroglucinol in human plasma by high-performance liquid chromatography-mass spectrometry |
| |
Authors: | Kim Hohyun Roh Hyeongjin Lee Hee Joo Chung Soo Youn Choi Sun Ok Lee Kyung Ryul Han Sang Beom |
| |
Affiliation: | Department of Pharmacokinetics, Seoul Medical Science Institute, Seoul Clinical Laboratories, 7-14, Dongbinggo-dong, Yongsan-gu, Seoul 140-809, South Korea. novakims@yahoo.com |
| |
Abstract: | A sensitive and selective liquid chromatographic method coupled with mass spectrometry (LC-MS) was developed for the quantification of phloroglucinol in human plasma. Resorcinol was used as internal standard, with plasma samples extracted using ethyl acetate. A centrifuged upper layer was then evaporated and reconstituted with mobile phase. The reconstituted samples were injected into a C(18) XTerra MS column (2.1 x 100 mm) with 3.5-microm particle size. The analytical column lasted for at least 500 injections. The mobile phase was 15% acetonitrile (pH 3.0), with flow-rate at 200 microl/min. The mass spectrometer was operated in negative ion mode with selective ion monitoring (SIM). Phloroglucinol was detected without severe interferences from plasma matrix when used negative ion mode. Phloroglucinol produced a parent molecule ([M-H](-)) at m/z 125 in negative ion mode. Detection of phloroglucinol in human plasma was accurate and precise, with quantification limit at 5 ng/ml. This method has been successfully applied to a study of phloroglucinol in human specimens. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|