首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Verification of accuracy and validity of gait phase detection system using motion sensors for applying walking assistive FES
Authors:Sunwoo Park  Kihong Ryu  Jungyoon Kim  Jongsang Son
Institution:Department of Biomedical Engineering , Institute of Medical Engineering, Yonsei University , Wonju , Gangwon , South Korea
Abstract:In this study, we have analysed heel strike (HS) and toe off (TO) of normal individuals and hemiplegic patients, taking advantage of output curves acquired from various sensors, and verified the validity of sensor detection methods and their effectiveness when they were used for hemiplegic gaits. Gait phase detections using three different motion sensors were valid, since they all had reliabilities more than 95%, when compared with foot velocity algorithm. Results showed that the tilt sensor and the gyrosensor could detect gait phase more accurately in normal individuals. Vertical acceleration could detect HS most accurately in hemiplegic patient group A. The gyrosensor could detect HS and TO most accurately in hemiplegic patient groups A and B. The detection of TO from all sensor signals was valid in both the patient groups A and B. However, the vertical acceleration detected HS validly in patient group A and the gyrosensor detected HS validly in patient group B.
Keywords:motion sensor  gait phase detection  walking assistive  functional electrical stimulation
设为首页 | 免责声明 | 关于勤云 | 加入收藏